Online Water Quality Monitor
Created with Pixso.

Dithizone Spectrophotometric Method Cadmium Water Quality Analyzer 0-5mg/L Range Online Heavy Metal Monitor with RS-232 4-20mA RS-485 Output

Dithizone Spectrophotometric Method Cadmium Water Quality Analyzer 0-5mg/L Range Online Heavy Metal Monitor with RS-232 4-20mA RS-485 Output

Brand Name: JUGE
Model Number: AG-TCd07
MOQ: 1 Unit
Price: USD2000-7000
Packaging Details: Neutral carton
Payment Terms: Online order,T/T
Detail Information
Place of Origin:
Nanjing, China
Certification:
IOS9001, CE
Type:
Cadmium
Principle:
Dithizone Spectrophotometric Method
Range:
0-5mg/L OEM
Accuracy:
≤±10%
Repeatability:
≤± 5%
Measurement Period:
30min
Output:
RS-232, 4-20mA, RS-485
Specification:
1500*500*450mm
Supply Ability:
100 Unit/Week
Highlight:

Dithizone Spectrophotometric Method Cadmium Analyzer

,

0-5mg/L Range Cadmium Water Quality Analyzer

,

RS-232 Output Online Heavy Metal Monitor

Product Description
Application

This method is applicable to wastewater with cadmium concentrations within the range of 0–5 mg/L.

Technical Parameters
Measurement Method Dithizone spectrophotometric method
Measurement Range 0–2 mg/L (OEM: 0–5 mg/L)
Accuracy ≤ ±10%
Repeatability ≤ ±5%
Measurement Cycle Minimum cycle is 30 minutes; color development time can be adjusted flexibly between 5–30 min based on environmental conditions.
Sampling Cycle Adjustable time interval (10–9999 min) or on-the-hour measurement mode.
Calibration Cycle Adjustable at any interval from 1–99 days.
Maintenance Cycle ≥ 168 hours, with each maintenance lasting approximately 1 hour.
Reagent Consumption < 0.50 CNY per sample.
Outputs RS-232, 4–20 mA, RS-485.
Environmental Requirements Temperature-controlled indoor environment (recommended: 5–40 °C); humidity 45%–85%.
Power Supply AC 220 ±10% V, 50 ±1% Hz, 5 A.
Dimensions Height 1500 × Width 500 × Depth 450 mm.
Additional Features
  • Data retention during abnormal alarms or power outages.
  • Touchscreen display and command input.
  • Automatic discharge of residual reactants and resumption of operation after power restoration or system reset.
Key Features

Unique Design: This product features a distinctive design that ensures lower failure rates, reduced maintenance requirements, minimal reagent consumption, and higher cost-effectiveness compared to similar products.

1. Selector Valve Assembly
  • Enables flexible reagent sampling timing and diverse channel configurations.
  • Minimizes dead volume for easy maintenance and extended service life.
2. Micro-Dosing Unit
  • Utilizes a visual photoelectric system for precise reagent measurement, eliminating quantitative errors caused by peristaltic pump tube wear.
  • Achieves accurate micro-dosing with 0.7 mL per dose, significantly reducing reagent usage.
3. Sample Introduction Unit
  • Employs a peristaltic pump with negative pressure suction, creating an air buffer zone between reagents and the pump tube to prevent corrosion.
  • Simplifies and enhances flexibility in reagent mixing.
4. Sealed Digestion Unit
  • Operates under high-temperature and high-pressure conditions to accelerate reactions.
  • Eliminates corrosion risks from volatile gases in open systems.
5. Reagent Tubing
  • Uses imported modified PTFE (polytetrafluoroethylene) transparent tubing with an inner diameter >1.5 mm to minimize clogging by particulate matter in water samples.
Analytical Principle

The aqueous sample undergoes hot-block digestion (150 °C ±5 °C) with concentrated HNO3 (1:1 v/v) followed by pH adjustment using ammonium citrate buffer. In alkaline medium (pH 10.0 ±0.1 maintained by NH4OH/NH4Cl buffer) with dithizone as the chromogenic agent, cadmium ions form a crimson-red Cd-dithizonate complex exhibiting maximum absorbance at 518 nm (ε = 8.5×104 L·mol−1·cm−1). A microprocessor-controlled portable photometer (e.g., Hach DR1900) quantifies cadmium concentration through dual-wavelength measurement (518 nm sample vs. 650 nm reference), with calibration curves validated per EPA Method 200.7 requirements (linear range: 5–200 μg/L, LOD 1.5 μg/L).